1. Standard mode fits :
- Working electrode (until OD: 6.0 mm)
- Reference electrode
- Platinum counter electrode 5.7 cm
- Purging tube
Set the Teflon cap adaptor to the Teflon cap.
The Teflon cap adaptor is fixed with a silicon O-ring to work in an oxygen-free condiction.
If you do not need the oxygen-free condiction, then the silicon O-ring can be taken out from the Teflon cap.
2. 10 mm WE mode fits :
- Working electrode (OD: 10.0 mm)
- Reference electrode
- Platinum counter electrode 5.7 cm
- Purging tube
Take out the Teflon cap adaptor from the Teflon cap.
Working electrode can be fitted directly to the Teflon cap.
The O-ring will fit perfectly. It permits to have the oxygen-free condition.
3. Small sample mode fits :
- Sample holder (OD: 9.0 mm)
→Working electrode
→Counter electrode
- Reference electrode
- Purging tube
Take out the Teflon cap adaptor from the Teflon cap, and set the Sample holder of the 9.0 mm of the diameter (optional item).
Set the working electrode and counter electrode into the Sample holder.
*Do not work in a oxygen-free condition.
4. Low temperature mode fits :
- Sample holder (OD: 9.0 mm)
→Reference electrode
- Platinum counter electrode 5.7 cm
- Micro electrode
- Purging tube
Take out the Teflon cap adaptor, and set the Sample holder of the 9.0 mm diameter (optional item).
Set the reference electrode into the sample holder.